7.8.10
Replace a valve module
Maintenance interval
Replace the valve modules when required, for example if a valve module is damaged.
Instruction
Follow the instruction below to replace a valve module.
WARNING
Disconnect power. Always disconnect power from the instrument
before replacing any component on the instrument, unless stated
otherwise in the user documentation.
Action
Step
Disconnect power from the instrument by switching off the Power switch.
The Power switch is located on the left hand side of the instrument.
1
Loosen the module with a Torx T20 screwdriver.
2
354
ÄKTA avant User Manual 29035184 AE
7 Maintenance
7.8 Replacement procedures
7.8.10 Replace a valve module