Possible cause and action
Problem
•
The air intake on the back or on the left side of the
instrument is covered
Make sure that none of the air intakes on the instru-
ment are covered.
•
Broken fans
Contact Service.
•
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35°C.
The temperature of the
instrument or an instru-
ment component is too
high
•
The gas spring is worn-out
Contact Service.
The Pump cover is
falling down
•
The waste tubing from the Buffer tray is loose
Contact Service.
Liquid from the Buffer
tray is flowing onto the
table
Power and ICU
Possible cause and action
Problem
•
The power cord is not connected
Connect the power cord to the wall outlet and to the
electrical inlet on the instrument. Make sure that the
cord is attached using the clip, thereby preventing the
cord from coming loose.
•
No electric current in the wall outlet
Make sure that there is electric current in the wall
outlet.
•
A fuse in the instrument ICU is broken
Contact Service.
•
The instrument is overheated
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact Service.
The instrument cannot
be turned on
ÄKTA avant User Manual 29035184 AE
407
8 Troubleshooting
8.7 Troubleshooting: Other components