Action
Step
Fit the syringe in the union Luer connector and inject the solution. DO NOT
remove the syringe.
3
Note:
Air trapped in the UV cell causes inaccurate measurements. To avoid introduc-
ing air into the UV cell, gently fill the union Luer up to the edge with test solu-
tion that is to be introduced, using the syringe. Then insert the syringe into
the union Luer.
Wait until the monitored absorbance value has stabilized.
4
In the Manual Instructions dialog box:
5
1
Select Monitors:Auto zero UV 2nd
2
Click Insert, then click Execute
Result: The UV absorbance is set to zero.
Remove the syringe.
6
Repeat the injections with the remaining four test solutions in increasing
concentration order. Use a new syringe for each solution.
7
After each injection, wait for a stable absorbance value. Note the measured
absorbance values for each solution.
8
ÄKTA avant User Manual 29035184 AE
333
7 Maintenance
7.7 Calibration procedures
7.7.4 Calibrate the UV monitors