Constraints on optional modules
The table below indicates usage constraints for the different external modules.
Constraints
External module
I/O-box E9 has no constraints.
I/O-box E9
I/O-box E9, 2nd requires I/O-box E9.
I/O-box E9, 2nd
Fraction collector 2 requires Fraction collector
and can not be installed at the same time as
Outlet valve 2nd or 3rd.
Fraction collector 2
External air sensor has no constraints
External air sensor
See respective section for each module.
Optional module mounted in Ex-
tension box
To optimize signal quality, the total cable length connecting all external modules
to the ÄKTA avant instrument should not exceed 10 m.
Note:
Edit System Properties
When a new module has been installed, the system properties have to be updated in
UNICORN. The system will restart automatically when the configuration has been changed
in the System Properties dialog box and then the system can be reconnected.
There are four main types of modules (named components in UNICORN) to select from:
•
Valves and pumps
•
Monitors and sensors
•
Fraction collectors
•
Other (e.g., I/O-box)
More than one component of the same type is only available if the first has been selected.
The selection made is reflected in which instructions and phase properties that are
available.
The following instruction gives a general description of how to update the system
properties in UNICORN.
ÄKTA avant User Manual 29035184 AE
107