
Action
Step
In the Process Picture:
3
•
Click the Inlet valve B icon.
•
Click the position of one of the inlets that will be used at the beginning
of the run.
Result: The inlet valve switches to the selected port.
In the Process Picture:
4
•
Click the System pumps icon.
•
Set Conc % B to 100% B and click Set % B.
•
Set the System flow to 1.0 ml/min for ÄKTA avant 25 or 5.0 ml/min for
ÄKTA avant 150.
•
Click Set flow rate.
Result: Only System Pump B is active, and a system flow through injection
valve waste starts.
ÄKTA avant User Manual 29035184 AE
169
5 Operation
5.4 Prime inlets and purge pump heads
5.4.1 Prime buffer inlets and purge system pumps
Содержание AKTA avant
Страница 1: ...ÄKTA avant User Manual ...
Страница 2: ...Page intentionally left blank ...
Страница 536: ...Page intentionally left blank ...
Страница 537: ...Page intentionally left blank ...