6-24
COBE 2991 Cell Processor
•
Essentials Guide
Processing and Common Operations
5
If predilution is not required, proceed to step 6. Or, to predilute the starting
cell product, perform the following steps:
a
Clamp the clear tubing between the RCD and the rotating seal
b
Place the cell product bag on the tabletop (Figure 6-17 on page 6-19).
c
Open the clamp on the blue-striped tubing.
d
Press the BLOOD IN button. Valve V1 opens, allowing processing/wash
solution to flow by gravity into the cell product bag. Manually agitate the
cell product bag to ensure adequate mixing (Figure 6-18 on page 6-20).
e
When the desired volume of processing/wash solution has entered the cell
product bag, press the STOP RESET button to close valve V1.
f
Close the slide clamp on the blue-striped tubing.
g
Allow the mixture to equilibrate (if necessary) for the appropriate time.
h
Remove the clamp from the tubing between the rotating seal and the
RCD.
6
Hang the cell product bag on the left hanger bar.
7
Open the slide clamp on the blue-striped tubing and let the starting cell
product flow by gravity into the cell processing bag.
After a short time, the starting cell product stops flowing. Depending on the
initial volume, some starting cell product may be left in the cell product bag.
Note:
DO NOT
press the AIR OUT
button for this
procedure.
8
Set the CENTRIFUGE SPEED control to 3000 rpm, then press the START
SPIN button. The centrifuge spins, pushing air and starting cell product back
up the blue-striped tubing to the cell product bag.
9
Press the STOP RESET button when air and starting cell product reach the
cell product bag. Additional starting cell product fills the cell processing bag.
Содержание COBE 2991
Страница 1: ......
Страница 18: ...This page intentionally left blank ...
Страница 19: ...This page intentionally left blank ...
Страница 20: ...This page intentionally left blank ...
Страница 22: ......
Страница 23: ......
Страница 24: ......
Страница 25: ......
Страница 26: ...Hướng dẫn sử dụng Part No 770273 010 Reorder No 709016 000 2005 05 Máy xử lý tế bào ...
Страница 27: ...COBE 2991 Cell Processor Essentials Guide 1 1 Introduction 1 1 Giới thiệu ...
Страница 37: ...Operator Controls 3 COBE 2991 Cell Processor Essentials Guide 3 1 Các kiểm soát của người vận hành ...
Страница 48: ...LED Programming 4 COBE 2991 Cell Processor Essentials Guide 4 1 4 Lên chương trình LED ...
Страница 59: ...Pin Diode Programming 5 COBE 2991 Cell Processor Essentials Guide 5 1 5 Lên chương trình Pin Diode ...
Страница 105: ...Troubleshooting 7 COBE 2991 Cell Processor Essentials Guide 7 1 7 Xử lý sự cố ...
Страница 116: ...General Maintenance 8 COBE 2991 Cell Processor Essentials Guide 8 1 8 Bảo dưỡng ...
Страница 120: ...Essentials Guide 1 6 17 05 10 29 08 AM ...
Страница 121: ...Essentials Guide Part No 770273 010 Reorder No 709016 000 2005 05 Cell Processor ...