MCTB
179
To begin calibration, send the low signal input to the loop control. If the process value does not match the
reference signal, adjust the low point offset so that the process value matches the reference input signal. Press
the button to switch to the high point offset and then apply the high signal input to the loop control. Adjust the
high point offset if necessary in order to make the process value match the reference signal.
Once the high point offset has been adjusted, press the button to switch to the low point offset and again apply
the low signal input to the loop control input. If the process value does not match the reference signal, adjust
the low point offset again so that the process value matches the reference input signal. Again, press the button
to return to the high point offset and apply the high signal input to the loop control. Adjust the high point offset
if necessary in order to get the process value to match the high reference signal. This process may have to be
repeated several times depending upon the amount of offset applied. Each time the low or high offset is
adjusted, it can affect the opposite end of the calibration curve.
If the known calibration error of the loop input is the same at both the minimum and maximum operating range
of the system, i.e., a linear offset, you can simply enter the difference between the reference signal and process
value for both the low point offset value and high point offset value. If both the low and high point offsets match,
it shifts the calibration curve thus applying the value of the low/high point offset equally across the entire range.
For example, if the range of operation is -50.0 to 350.0 degrees and the error between the reference signal and
the process value is 0.7 at both the low and high range resulting in process values of -49.3 and 350.7, simply
enter the value of -0.7 for both the low and high point offset. This will shift the curve equally at both ends of the
curve so the resulting process values will be -50.0 and 350.0.
Once the calibration offset is complete, press the “Done” button to return to the main Calibration screen.
10.5.2 Limit Calibration Offset
The calibration offset provides a single point shift. If the limit control is used for high/low operation and the error
differs from one end of the calibration curve to the other, the shift value can be used to split the difference
between the two. This will make the error equal at both ends of the curve or it can be used to shift it to reduce
error at one end of the curve; however, it will also increase the error at the opposite end of the curve.
Содержание MCTB 4.3
Страница 24: ...MCTB 24...
Страница 195: ...MCTB 195...
Страница 227: ...MCTB 227 A 5 SmartIO Monitor Module Specifications...
Страница 228: ...MCTB 228 A 6 Expansion IO Module Specifications...
Страница 229: ...MCTB 229 A 7 Power Supply Specifications DIN Rail Mount Power Supply Specifications...
Страница 230: ...MCTB 230...
Страница 231: ...MCTB 231 Open Frame Power Supply Specifications...
Страница 232: ...MCTB 232...
Страница 233: ...MCTB 233 A 8 Power On Delay Relay Specifications GE1A C10HA110...
Страница 234: ...MCTB 234...