3 - 1
INP-TN4PHAV-E
This unit is designed to be panel mounted.
3.1
Mounting location
Select the following location for mounting the unit.
(1) A place that is not subject to vibration or impact.
(2) A place where there is no corrosive gas.
(3) A place that is subject to little temperature variation and is
close to normal temperature (23
°
C)
(4) A place that is not struck directly by strong radiant heat.
(5) As humidity affects the ink and recording paper, select a place
that is in the range 45 to 80% RH.
(6) Mount the unit horizontally, with no tilt to the left or right.
(The forward tilt should be 0 (but the unit may be inclined 0 to
30
°
rearwards.)
3.2
External dimensions and panel cut out dimensions (unit: mm)
3. MOUNTING METHOD
∠ α
∠ α
= 60 to 90
280
312
288
288
360MIN
281
+2
0
281
+2
0
320MIN
280
160
39
18
199
Transmission terminal
(option)
Alarm unit (option)
(Alarms 7 to 12)
Alarm, external
control unit
(option) (Alarms 1 to 6)
Mounting
attachment
2
≤
t
≤
30
Panel
Power supply terminal
Input terminals
(inputs 7 to 12)
Input terminals
(inputs 1 to 6)
Panel cut dimensions (unit: mm)
Mass: Approximately 6kg (without options)
Approximately 7kg (with all options)
Power consumption: Approximately 22VA
(100V AC without options)
Approximately 37VA
(100V AC with all options)