7 - 5
INP-TN4PHAV-E
7.4
Setting the chart speed (main chart speed/sub-chart speed)
Explanation
• Main chart speed: This is the procedure for setting the chart speed in normal operation.
The setting range is 5 to 1500mm/h. (Can be set in 1mm/h steps.)
• If the case of a continuous recording type, if the chart speed is too fast, the result is dashed
line recording instead of continuous recording. (As a general criterion, 300mm/h or more)
• Please note that the following digital print-outs are not possible if the chart speed of continuous
recording type is 301mm/h or more, that of dot recording type is 51mm/h or more.
'Periodic print-out' 'Scale print-out' 'Message print-out' 'Alarm print-out' 'Parameter print-out' 'Ink
Out print-out'.
However, a 'Scale print-out' 'Message print-out' can be made manually.
See Section 7.11.
• The sample time in continuous recording varies depending on the chart speed
450
Sample time =
Chart speed (mm/h)
(But not faster than 3 seconds.)
Example:
Chart speed (mm/h)
10
20
25
50
100
150
Sample time
45
Repeated 22, 23
20
9
Repeated 4, 5
3
Key actuation
Changing the normal chart speed of 25mm/h to 20mm/h
(example)
Key operation
Explanation
Display
SEL
Press the SEL key twice to display the main
chart speed. (In case where pass code =0)
Press the
key to set to '20'
ENT
Press the ENT key to effect registration.
There is a move the display of the next parameter.
MAIN CHART SPEED
25mm/h
MAIN CHART SPEED
20mm/h
SUB CHART SPEED
25mm/h