7 - 12
INP-TN4PHAV-E
7.7
Setting record ranges
Explanation
An individual record range is set for each channel. 'Record range' means the 0% and 100% position
scale on recording paper. If DC voltage input scaling is set and used, set the record range after setting
scaling. If scaling is set, the decimal point position is the same one as set by the engineering value.
Key actuation
The record range of channel 1 is changed from 0 to 100
°
C to -50 to 50
°
C.
(example)
Key operation
Explanation
Display
SEL
Press the SEL key several times to give the
recording range display.
ENT
Since channel No. 1 is selected, press the ENT
key.
Press the
key to change the range lower limit
ENT
from "0" to "-50" and press the ENT key.
Press the
key to change "100" to "50" and
ENT
press the ENT key.
(When the ENT key is pressed, the channel No.
flashes and the setting is completed.
Follow the same procedure for settings in other
channels.)
ch 1 RANGE
0 to 100
°
C
ch 1 RANGE
0 to 100
°
C
ch 1 RANGE
-50 to 100
°
C
ch 1 RANGE
-50 to 100
°
C
Table 1 Specifiable ranges record ranges
Specifiable ranges
370.0 to 1790.0
°
C
-30.0 to 1790.0
°
C
-30.0 to 1790.0
°
C
-230.0 to 1400.0
°
C
-230.0 to
830.0
°
C
-230.0 to 1130.0
°
C
-230.0 to
430.0
°
C
-30.0 to 1330.0
°
C
-30.0 to 1790.0
°
C
-230.0 to
930.0
°
C
-230.0 to
430.0
°
C
-30.0 to 1330.0
°
C
-230.0 to
630.0
°
C
-230.0 to
630.0
°
C
-55.0 to +55.0mV
-550.0 to +550.0mV
-5,500 to +5,500V
55.00 to +55.00V
Type
Thermocouples
B
R
S
K
E
J
T
N
W
L
U
PN
Resistance
JPt 100
bulbs
Pt100
DC voltage
Reference range
400 to 1760
°
C
0 to 1760
°
C
0 to 1760
°
C
-200 to 1370
°
C
-200 to
800
°
C
-200 to 1100
°
C
-200 to
400
°
C
0 to 1300
°
C
0 to 1760
°
C
-200 to
900
°
C
-200 to
400
°
C
0 to 1300
°
C
-200 to
600
°
C
-200 to
600
°
C
-50 to +50mV
-500 to +500mV
-5 to
+5V
-50 to +50V
Reference range
725 to 3200
°
F
32 to 3200
°
F
32 to 3200
°
F
-328 to 2498
°
F
-328 to 1472
°
F
-328 to 2012
°
F
-328 to
752
°
F
32 to 2372
°
F
32 to 3200
°
F
-328 to 1652
°
F
-328 to
752
°
F
32 to 2372
°
F
-328 to 1112
°
F
-328 to 1112
°
F
for record range
698.0 to 3254.0
°
F
-22.0 to 3254.0
°
F
-22.0 to 3254.0
°
F
-382.0 to 2552.0
°
F
-382.0 to 1526.0
°
F
-382.0 to 2066.0
°
F
-382.0 to
806.0
°
F
-22.0 to 2426.0
°
F
-22.0 to 3254.0
°
F
-382.0 to 1706.0
°
F
-382.0 to
806.0
°
F
-22.0 to 2426.0
°
F
-382.0 to 1166.0
°
F
-382.0 to 1166.0
°
F
The range -32767 to 32767 (decimal points can be located wherever required) can be specified for the record range
in scaling or difference calculation specifications.