6 - 7
INP-TN4PHAV-E
6.10 Display when data backup batteries need to be replaced
When the voltage of back-up batteries becomes low, a display indicating that they need to be replaced
is given.
'Battery End' is displayed in the display section.
Example
Ch6
123.5
°
C
Battery End
Replace the batteries promptly when a 'Battery End' display appears page 8-3.
6.11 Display of fault in recording head carriage
If a fault in the recording head carriage occurs and the recording head can no longer function normally,
a fault display is given and the recording operation stops.
Example
Ch6
123.5
°
C
Carriage Alarm
If a 'Carriage Alarm' display is given, check the following points.
(1) Is foreign matter adhering to the recording head carrier shaft ?
(2) Is the wire that drives the recording head broken or slack ?
(3) Has the recording paper lifted up and come into contact with the recording head ?
(4) Is the recording head set in place correctly ?
After eliminating the cause of the fault, switch on the power supply of the main unit.
6.12 Order of priority of state displays
If the items noted below occur simultaneously, the corresponding displays are given in the indicated order.
1.
Chart end
2.
Carriage alarm
3.
Ink end
4.
Battery end
5.
Alarm
Note: When the state displays 1 and 2 above are given, the SELECT key is inoperative.
However, the DISPLAY and FEED keys are operative.