12 - 2
INP-TN4PHAV-E
Recording section
• Recording system:
Ink jet system, 6 colours
• Effective recording width:
180mm
• Recording colours:
1st, 7th (orange), 2nd, 8th (green), 3rd, 9th (purple), 4th, 10th (red),
5th, 11th (black), 6th, 12th (blue)
• Chart paper:
folding, total length 20m
• Chart speed:
5 to 300mm/h continuous record (300mm/h is the general standard)
301 to 1500mm/h discontinuous records
Dot record type ......................... 5 to 1500mm/h
All settable in 1mm/h steps.
• Speed setting method:
Set from keyboard.
• Sample time:
Dot records ............................... 30 seconds/for all channels
Continuous records ................... Depends on chart speed.
Calculation formula:
450
sample time (seconds) =
chart speed (mm/h)
But is not faster than 3 seconds.
• Measurement period:
1 to 3 input points ..................... 160ms
6 to 12 input points ................... 320ms
• Ink life (depends on conditions): approximately 6 months for 6 point continuous records at a record-
ing paper feed speed of 25mm/h
• Accuracy, resolution: Performance at standard conditions (23
±
2
°
C, 55
±
10% RH, power supply volt-
age and frequency fluctuation within
±
1%, warm-up time 30 minutes or more,
vertical mounting, environment with no adverse effects of external noise, etc.)
Type of input
Thermocouple
B
R
S
K
E
J
T
N
W
L
U
PN
Resistance
JPt100
bulb
Pt100
DC voltage
-50 to +50mV
-500 to +500mV
-5 to +5V
-50 to +50V
play)
Resolution
0.1
°
C
0.1
°
C
0.1
°
C
0.1
°
C
0.1
°
C
0.1
°
C
0.1
°
C
0.1
°
C
0.1
°
C
0.1
°
C
0.1
°
C
0.1
°
C
0.1
°
C
10
µ
V
100
µ
V
1mV
10mV
Indication (digital dis
Accuracy
±
(0.15% +1 digit)
(Does not include reference
junction compensation error.)
±
(0.15% +1 digit)
±
(0.15% +1 digit)
Record
Accuracy
Indication
precision
±
(0.25%
record span)
Resolution
0.1mm
Note 1: The rating of indication accuracy is shown in % within the input span.
Note 2: Indication accuracy at 400 to 600
°
C of B-thermocouple is
±