INP-TN4PHAV-E
Step 4
Remove the connectors (2 places) connecting the
display unit to the main unit.
Step 5
Loosen the mounting brackets on both sides of
the fluorescent lamp at the rear of the display unit.
Remove the brackets from the display unit.
Step 6
Remove the lamp from the mounting brackets,
then pull off the connector by pushing up the tab
on the connector between the lamp and the in-
verter.
Step 7
• Set a new lamp (with cable and connector).
It should be set in the order of Step 6
→
Step 5
→
Step 4
→
Step 3
→
Step 2
→
Step 1 .
Step 8
• Turn ON the power and check that the lamp lights up.
• The replacement lamp (with cable and connector) should be ordered specifying the follow-
ing Part No.
Part No. PHZL8001
Reference: Life of lamp
• About 1 year (continuous)
Connector
Connector
Screw
Lamp mounting
bracket
Connector