1 - 40
CHARACTERISTICS
CHARACTERISTICS
SPECIFICATION
ADDITIONAL
INFORMATION
The PM3390B and PM3394B, including standard accessories, conform with the
EEC Directive 89/336 for EMI immunity, as defined by EN50081-1 (1992) and
EN50082-1 (1992), with the addition of the following tables.
Table 5.
Susceptibility: no visible disturbance
Frequency range:
10 kHz .. 25 MHz
E = < 0.1V/m
E = 1 V/m
E = 3 V/m
Stand alone
2 mV/div ... 5 V/div
2 mV/div ... 5 V/div
2 mV/div ... 5 V/div
With PM9020/09x
2 mV/div ... 5 V/div
20 mV/div ... 5 V/div
50 mV/div ... 5 V/div
Table 6.
Susceptibility: no visible disturbance
Frequency range:
25 MHz ... 1 GHz
E = < 0.1V/m
E = 1 V/m
E = 3 V/m
Stand alone
2 mV/div ... 5 V/div
2 mV/div ... 5 V/div
10 mV/div ... 5 V/div
With PM9020/09x
2 mV/div ... 5 V/div
100 mV/div ... 5 V/div
200 mV/div ... 5 V/div
Table 7.
Susceptibility: disturbance less than 10 % of full scale
Frequency range:
10 kHz.. 25 MHz
E = < 0.1V/m
E = 1V/m
E = 3 V/m
Stand alone
N/A
N/A
N/A
With PM9020/09x
N/A
5 mV/div ... 10 mV/div
10 mV/div ... 20 mV/div
Table 8.
Susceptibility: disturbance less than 10 % of full scale
Frequency range:
25 MHz ... 1 GHz
E = < 0.1V/m
E = 1V/m
E = 3 V/m
Stand alone
N/A
N/A
5 mV/div
With PM9020/09x
N/A
20 mV/div ... 50 mV/div
50 mV/div ... 100 mV/div
For conditions not specified in tables 5-8,, a susceptibility effect of more than 10 % is possible.