PERFORMANCE TEST
4 - 39
Test set-up:
Settings/procedure and requirements:
1
-
Apply a 100 MHz sine-wave signal of 1 V(pp into 50
Ω
) from the SG 503 to
input CH1.
2
-
Press the AUTOSET key and set CH1 to 0.5V/div.
3
-
Use a 50
Ω
termination. For instruments with switchable 50
Ω
input
impedance it is recommended to make use of this feature (via
VERT MENU key).
4
-
Set the input coupling of CH1 to DC and POSition the signal in the vertical
center of screen.
5
-
Select ’trig’ from in the menu under the TB MODE mode key.
6
-
Press the TRIGGER menu key and select level-pp ’off’ and ’dc’ trigger
coupling from the TRIGGER MAIN TB menu
7
-
Adjust TRIGGER LEVEL for a correctly triggered display.
8
-
Decrease the amplitude of the input signal.
9
-
Verify that the signal is well-triggered at amplitudes of 0.6 divisions and more.
10 -
Decrease the input frequency to 50 kHz.
11 -
Verify that the signal stays well-triggered at amplitudes of 0.6 divisions and
more.
12 -
Increase the input frequency to 200 MHz.
13 -
Increase the input voltage to 1.2 division.
14 -
Turn TRIGGER LEVEL.
15 -
Verify that the signal is well-triggered at amplitudes of 1 division and more.
16 -
Apply a 300 MHz sine-wave signal of 2V (pp into 50
Ω
) from the SG 504 to
input CH1.
17 -
Adjust the input voltage to 2 divisions. Signal must be in vertical center of
screen.
18 -
Verify that the signal is well-triggered at amplitudes of 2 divisions and
more; adjust TRIGGER LEVEL when necessary.