PERFORMANCE TEST
4 - 43
Test equipment:
Fluke 5500A mode: levsine
(Alternative: constant amplitude sine wave generators SG 503 + SG 504).
Test set-up:
Settings/procedure and requirements:
1
-
Apply a 100 MHz sine-wave signal of 1 V(pp into 50
Ω
) from the SG 503 to
input CH1.
2
-
Press the AUTOSET key and set CH1 to 0.5V/div.
3
-
Use a 50
Ω
termination. For instruments with switchable 50
Ω
input
impedance it is recommended to make use of this feature (via
VERT MENU KEY).
4
-
Set the input coupling of CH1 to DC and POSition the signal in the vertical
center of screen.
5
-
Select ’trig’ from the menu under the TB MODE menu key.
6
-
Press the TRIGGER menu key and select level-pp ’off’ and ’dc’ trigger
coupling from the TRIGGER MAIN TB menu.
7
-
Adjust TRIGGER LEVEL for a correctly triggered display.
8
-
Press the DTB menu key and select DEL’D TB ’on’ and MAIN TB ’on’ in
the related menu.
9
-
Set MAIN TB to 200 ns/division and DELAYED TIMEBASE to
20.0 ns/division.
10 -
Adjust the DELAY control in the DELAYED TIMEBASE section to a delay
time of 1.000
µ
s.
11 -
Select ’trig’d’ and ’dc’ coupling from the DELAYED TIME BASE menu, and
press the front panel key TRIG1. Or TRIG2 (if CH2 on), TRIG3 (if CH3 on),
TRIG4 (if CH4 on).
12 -
Adjust the DEL’D TB trigger level via the
∆
control for a well- triggered
signal (intensified part must be visible).
13 -
Operate the TRACK control to separate MAIN TB and DEL’D TB for clearly
visible displays.