4 - 52
PERFORMANCE TEST
Settings/procedure:
1
-
Adjust the generator to a 1 kHz sine-wave of 2 V(pp) applied to CH1.
2
-
Press the AUTOSET key.
3
-
Press the AUTO RANGE key of the main time base MAIN TB.
4
-
Vary the generator output frequency between 1kHz and maximum
(10 MHz approxinately).
Requirement:
-
Check that between 2 and 6 waveform periods are displayed.
4.3.44
Testing the optional auxiliary outputs
MTB-GATE and DTB-GATE outputs
Test equipment:
measuring oscilloscope.
Test set up:
Settings/procedure:
Oscilloscope under test:
-
Take care that no input signal is applied to the oscilloscope inputs.
-
Press the STATUS and TEXT OFF keys simultaneously. This assures that the
oscilloscope occupies its default position.
-
Press the DTB menu key to display the DELAYED TIME BASE menu. Select
the ‘on’ position of the DEL’D TB softkey. The result is that MTB is in 1.00 ms/
div and that DTB is in 100
µ
s/div and that both time bases are on.
-
Apply a 50
Ω
coaxial cable to rear panel output ‘MTB-GATE’.
-
Apply another 50
Ω
coaxial cable to rear panel output ‘DTB-GATE’.