1 - 6
CHARACTERISTICS
CHARACTERISTICS
SPECIFICATION
ADDITIONAL
INFORMATION
1.1.7
Dynamic Range
PM3390B/PM3394B
Up to 50 MHz
±
12 div
Symmetrical
Up to 200 MHz
±
4 div
Symmetrical
PM3380B/PM3384B
Up to 25 MHz
±
12 div
Symmetrical
Up to 100 MHz
±
4 div
Symmetrical
PM3370B
Up to 15 MHz
±
12 div
Symmetrical
Up to 60 MHz
±
4 div
Symmetrical
1.1.8
Position Range
POSITION RANGE
+ or -
≥
8 div
Symmetrical
1.1.9
Trace Separation
TRACE SEPARATION
MTB and DTB
Min. range
+ or -
≥
4 div
MTB fixed, DTB shifts
1.1.10
Max. Input Voltage
MAX. INPUT VOLTAGE
See Note 1
In high Z position
According to EN61010-1 Cat II Pol 2.
Max. input voltage
150 V rms
See Note 2.
In 50
Ω
position
dc
±
5V
ac rms
5V
See Note 3
ac peak
±
50V
See Note 3
Note 1: The instrument should be properly grounded through the protective
ground conductor of the power cord.
Note 2: Up to 20 KHz; >20 kHz see figure 1.1
Note 3: Maximum of 50 mJ during any 100 ms interval.