PERFORMANCE TEST
4 - 27
Test set-up:
Settings/procedure:
1
-
Apply a 2 kHz sine-wave signal of 800 mV (pp) to input CH1.
2
-
Press the AUTOSET key and set CH1 to 0.1V/div.
3
-
Adjust the generator signal to a trace height of 8 divisions.
4
-
Press the DISPLAY menu key.
5
-
Press the X-DEFL softkey in the DISPLAY menu.
6
-
Select ’on’ and ’ch1’ as X-SOURCE in the X-DEFL menu.
7
-
Use the CH1 POS and X POS controls to obtain the display shown in the
figure below.
Requirements:
-
Verify that a line with an angle of 45
°
is displayed.
MAT3837
Figure 4.4
X deflection