CHARACTERISTICS
1 - 39
CHARACTERISTICS
SPECIFICATION
ADDITIONAL
INFORMATION
The PM3370B, PM3380B and PM3384B, including standard accessories,
conform with the EEC Directive 89/336 for EMI immunity, as defined by EN50081-
1 (1992) and EN50082-1 (1992), with the addition of the following tables.
Table 1.
Susceptibility: no visible disturbance
Frequency range:
10 kHz .. 25 MHz
E = < 0.1V/m
E = 1 V/m
E = 3 V/m
Stand alone
2 mV/div ... 5 V/div
2 mV/div ... 5 V/div
2 mV/div ... 5 V/div
With PM9010/09x
2 mV/div ... 5 V/div
20 mV/div ... 5 V/div
100 mV/div ... 5 V/div
Table 2.
Susceptibility: no visible disturbance
Frequency range:
25 MHz ... 1 GHz
E = < 0.1V/m
E = 1 V/m
E = 3 V/m
Stand alone
2 mV/div ... 5 V/div
2 mV/div ... 5 V/div
10 mV/div ... 5 V/div
With PM9010/09x
2 mV/div ... 5 V/div
100 mV/div ... 5 V/div
500 mV/div ... 5 V/div
Table 3.
Susceptibility: disturbance less than 10 % of full scale
Frequency range:
10 kHz .. 25 MHz
E = < 0.1V/m
E = 1V/m
E = 3 V/m
Stand alone
N/A
N/A
N/A
With PM9010/09x
N/A
5 mV/div ... 10 mV/div
20 mV/div ... 50mV/div
Table 4.
Susceptibility: disturbance less than 10 % of full scale
Frequency range:
25 MHz ... 1 GHz
E = < 0.1V/m
E = 1V/m
E = 3 V/m
Stand alone
N/A
N/A
2 mV/div ... 5 mV/div
With PM9010/09x
N/A
20 mV/div ... 50 mV/div
50 mV/div .... 200 mV/div
For conditions not specified in tables 1 - 4, a susceptibility effect of more than 10 % is possible.