1 - 34
CHARACTERISTICS
CHARACTERISTICS
SPECIFICATION
ADDITIONAL
INFORMATION
1.14 AUTO SET & CALIBRATION
1.14.1
Auto Set
Vertical deflection
2...5 div
Note 1
Horizontal deflection
Max. 6 periods on CRT
Note 1
at input signal
10 mV...25V
40 Hz...30 MHz
Note 1: AUTO SET selects the proper channel, sets vertical deflection, timebase
speed, intensity, and triggering for an easy-to-read display of input
signals, or the user programble AUTO SET items.
1.14.2
Calibration
CALIBRATION
FACILITIES
Auto cal
See Note 1
Note 1: Calibrates vertical offset and gain, horizontal offset and gain and sweep
time, trigger offset and gain.
1.15 POWER SUPPLY AND BATTERY BACKUP
1.15.1
Power Supply
LINE VOLTAGE
ac (rms)
Operation
100V to 240V
CAT II Pol 2
Tolerance
±
10%
LINE FREQUENCY
Nominal
50 Hz to 400 Hz
Limits of operation
45 Hz to 440 Hz
LINE WAVEFORM
At nominal source
Max. waveform
voltage
deviation factor
10%
Crest factor
1.27 to 1.56