19 Thermal energy rate measurement
FLUXUS F502TE
19.2 Transmitter settings
139
UMFLUXUS_F502TEV1-5EN, 2020-05-29
19.2.2
Displaying the transmitter terminals
• Select the menu item
Special Funct.\SYSTEM settings\Proc. inputs\Terminal
.
• Press ENTER.
The terminals for the connection of the temperature input
T1
are displayed.
• Press ENTER.
The terminals for the connection of the temperature input
T2
are displayed.
• Press ENTER.
19.2.3
Selection of the temperature probe
• Select the menu item
Special Funct.\SYSTEM settings\Proc. inputs
.
• Select
PT100/PT1000
from the scroll list.
• Press ENTER.
• Select the temperature probe for the input
T1
.
• Press ENTER.
• Select the temperature probe for the input
T2
.
• Press ENTER.
19.2.4
Activation of the temperature inputs
The temperature inputs are displayed one after another in the program branch
Output Options
.
• Select
yes
if the temperature input
T1
is to be activated.
• Press ENTER.
This display will not be indicated if
Thermal energy
has been selected as physical quantity.
• Select
yes
if the temperature input
T2
is to be activated.
• Press ENTER.
This display will not be indicated if
Thermal energy
has been selected as physical quantity.
Special Funct.\SYSTEM settings\Proc. inputs\Terminal
Input T1
Input T2
Special Funct.\SYSTEM settings\Proc. inputs
Input T1
Input T2
Output Options\...\Temperature T1
Output Options\...\Temperature T2
Notice!
If a temperature input is activated, the total number of measured values that can be stored will be reduced.
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......