19 Thermal energy rate measurement
19.2 Transmitter settings
FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
138
19.2
Transmitter settings
If the position of the transducers and the temperature inputs are changed, the following settings have to be made.
19.2.1
Assignment of the transducers and the temperature inputs
The position of the transducers and the temperature inputs can be assigned in accordance with the application.
• Select the menu item
Special Funct.\SYSTEM settings\Proc. inputs\Link temperature
.
• Press ENTER.
• Select
heat
for a heating application or
chill
for a cooling application.
• Press ENTER.
• Select
return
if the transducers are mounted on the return line or
supply
if they are mounted on the supply line.
• Press ENTER.
• Select
sign
if the sign of the thermal energy rate is to be considered or
absolute
if just the absolute value of the thermal
energy rate is to be displayed.
• Press ENTER.
• Select
yes
if the aggregate state of the fluid changes between the return and supply line. Select
no
if it does not change.
• Press ENTER.
• Select the temperature input to be assigned to the supply temperature.
• Press ENTER.
• Select the temperature input to be assigned to the return temperature.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. inputs\Link temperature
Special Funct.\...\Thermal energy
Special Funct.\...\Transd. Location
Special Funct.\...\Thermal energy
Special Funct.\...\With PhaseChange
Special Funct.\...\T-Supply\Input T1
Special Funct.\...\T-Return\Input T2
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......