15 Data transmission
FLUXUS F502TE
15.3 Transmission parameters
107
UMFLUXUS_F502TEV1-5EN, 2020-05-29
15.3
Transmission parameters
• the transmitter sends CRLF-terminated ASCII
• max. line length: 255 digits
RS232
default: 9600 bits/s, 8 data bits, even parity, 2 stop bits, RTS/CTS protocol (hardware handshake)
The transmission parameters of the service interface RS232 can be changed.
• Enter the HotCode
232-0-
(see section 8.4).
• Set the transmission parameters in the 4 scroll lists.
• Press ENTER.
–
baud
: baud rate
–
data
: number of data bits
–
par
: parity
–
st
: number of stop bits
RS485
default: 9600 bit/s, 8 data bits, even parity, 1 stop bit
• Select the menu item
Special Funct.\SYSTEM settings\Network
to change the settings of the transmission
parameters.
• Press ENTER.
This display will only be indicated if the transmitter has an interface RS485.
• Press ENTER to confirm the device address in the network.
• Select
default
to display the default transmission parameters.
• Select
setup
to change the transmission parameters.
• Press ENTER.
• Set the transmission parameters in the 3 scroll lists.
• Press ENTER.
–
baud
: baud rate
–
parity
: parity
–
st
: number of stop bits
The default transmission parameters will be used if
default
is selected and the transmission parameters have not been
changed.
Special Funct.\SYSTEM settings\Network
Special Funct.\SYSTEM settings\Network\Device address
Special Funct.\SYSTEM settings\Network\RS485 protocol
BAUD<data par st
9600 8bit EVEN 2
BAUD< parity st
9600 EVEN
2
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......