16 Advanced functions
FLUXUS F502TE
16.4 Profile correction
113
UMFLUXUS_F502TEV1-5EN, 2020-05-29
16.4
Profile correction
The following settings can be made for the calculation of the fluid mechanics calibration factor k
Re
:
•
off
: profile correction 1.0
•
on
: profile correction 2.0 at ideal inflow conditions (default)
•
With disturbance
: profile correction 2.0 at non ideal inflow conditions
The following steps are necessary to set the profile correction:
• Selection of the profile correction setting for all measuring channels in the program branch
Special Funct.
• Input of the disturbance distance in the program branch
Parameter
if
With disturbance
has been selected
If
With disturbance
has been selected, the transducers have to be mounted in reflection arrangement, X arrangement
or displaced X arrangement to compensate transverse flow effects. When mounting in X arrangement, it is essential to set
the same parameters for both measuring channels and to activate for them a calculation channel with average generation.
Selection of the setting
• Select the menu item
Special Funct.
in the program branch
Measuring
.
• Press ENTER until the menu item
ProfileCorr 2.0
is displayed.
• Select a list item (default:
on
).
• Press ENTER.
Input of the disturbance distance
If
With disturbance
is selected in the menu item
Special Funct.\SYSTEM settings\Measuring\
ProfileCorr 2.0
, the disturbance distance has to be entered in the program branch
Parameter
.
• Enter the disturbance distance.
• Press ENTER.
Measurement
When starting the measurement, it is checked whether the measurement arrangement is appropriate.
16.5
Uncorrected flow velocity
For special applications, the uncorrected flow velocity might be of interest.
The profile correction for the flow velocity is activated in the menu item
Special Funct.\SYSTEM settings\
Measuring\Flow Velocity
.
• Select
normal
to display and output the flow velocity with profile correction.
• Select
uncorr.
to display and output the flow velocity without profile correction.
• Press ENTER.
If
uncorr.
is selected, it has to be confirmed each time the program branch
Measuring
is selected whether the profile
correction is to be used.
If
no
is selected, the profile correction will be switched off. All physical quantities will be calculated with the uncorrected
flow velocity.
Special Funct.\...\Measuring\ProfileCorr 2.0
Special Funct.\SYSTEM settings\Measuring\Flow Velocity
Disturb.distance
2.3
m
PROFILE CORR.
>NO<
yes
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...