14 Data logger
14.3 Configuration of the data logger
FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
102
Starting time
It is possible to set a starting time if it is necessary to synchronize the storing of measured values for several transmitters.
• Select the moment at which the storing has to start.
Ringbuffer
The setting of the ringbuffer influences the storing of measured values as soon as the data logger is full:
• If the ringbuffer is activated, the data logger capacity will be halved. The oldest measured values will be overwritten. Only
the data logger memory that was free during the activation will be used by the ringbuffer. If more memory is required, the
data logger should be previously cleared.
• If the ringbuffer is deactivated, the storing of measured values will be stopped.
• Select
ON
to activate the ringbuffer.
• Press ENTER.
Storage mode
• Select the storage mode.
• Press ENTER.
If
sample
is selected, the current measured value will be used for the storing and online transmission of data.
If
average
is selected, the average of all undamped measured values of a storage interval will be used for the storing and
online transmission of data.
Special Funct.\SYSTEM settings\Storing\Start logger
display
description
Promptly
The storing starts immediately.
On full 5 min.
The storing starts in the next full 5 minutes.
On full 10 min.
The storing starts in the next full 10 minutes.
On quarter hour
The storing starts in the next full 15 minutes.
On half hour
The storing starts in the next full 30 minutes.
On full hour
The storing starts in the next full 60 minutes.
Example
actual time: 09:06 am
setting:
On full 10 min.
The storing starts at 09:10 am.
Special Funct.\SYSTEM settings\Storing\Ringbuffer
Special Funct.\SYSTEM settings\Storing\Storage mode
Notice!
The storage mode does not affect the outputs.
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......