11 Maintenance and cleaning
11.2 Cleaning
FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
96
11.2
Cleaning
Stainless steel housing
• Clean the housing with a soft cloth and care and cleaning spray for stainless steel.
Aluminum housing
• Clean the housing with a soft cloth. Do not use detergents.
Transducers
• Remove traces of coupling compound from the transducers with a soft paper towel.
11.3
Calibration
If installed as recommended in an appropriate location, used cautiously and serviced conscientiously, no troubles should
appear.
The transmitter has been calibrated at factory and, usually, a recalibration of the transmitter is not necessary.
A recalibration is recommended if:
• the contact surface of the transducers show visible wear or
• the transducers were used for a prolonged period at high temperatures (several months > 130 °C for normal transducers
or > 200 °C for high temperature transducers)
In order to realize a recalibration under reference conditions, either the transmitter, the transducers or both have to be sent
to FLEXIM.