16 Advanced functions
16.3 Cut-off flow
FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
112
16.3
Cut-off flow
The cut-off flow is a lower limit for the flow velocity. All measured flow velocities that are lower than the limit and their
derived quantities are set to zero.
The cut-off flow can depend on the flow direction. The cut-off flow is set in the menu item
Special Funct.\
SYSTEM settings\Measuring\Cut-off Flow
.
• Select
sign
to define a cut-off flow depending on the flow direction. One limit is set for the positive and another for the
negative flow velocity.
• Select
absolut
to define a cut-off flow not depending on the flow direction. A limit is set for the absolute value of the
flow velocity.
• Press ENTER.
• Select
factory
to use the default limit of 2.5 cm/s (0.025 m/s) for the cut-off flow.
• Select
user
to enter the cut-off flow.
• Press ENTER.
If
Cut-off Flow\sign
and
user
are selected, 2 values have to be entered:
• Enter the cut-off flow.
• Press ENTER.
All positive values of the flow velocity below this limit are set to zero.
• Enter the cut-off flow.
• Press ENTER.
All negative values of the flow velocity above this limit will be set to zero.
If
Cut-off Flow\absolut
and
user
are selected, only one value has to be entered:
• Enter the cut-off flow.
• Press ENTER.
All absolute values of the flow velocity below this limit are set to zero.
Special Funct.\SYSTEM settings\Measuring\Cut-off Flow
Special Funct.\...\+Cut-off Flow
Special Funct.\...\-Cut-off Flow
Special Funct.\...\Cut-off Flow
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......