3 General principles
FLUXUS F502TE
3.4 Undisturbed flow profile
17
UMFLUXUS_F502TEV1-5EN, 2020-05-29
disturbance source: valve
inlet: l ≥ 40 D
outlet: l ≥ 5 D
disturbance source: reducer
inlet: l ≥ 10 D
outlet: l ≥ 5 D
disturbance source: pump
inlet: l ≥ 20 D
Tab. 3.1:
Recommended distance from disturbance sources
D – nominal pipe diameter at the measuring point
l – recommended distance between disturbance source and transducer position
l
l
l
l
l