4 Product description
4.1 Measuring system
FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
18
4
Product description
4.1
Measuring system
The measurement system consists of the transmitter, the ultrasonic transducers, the temperature probes and the pipe on
which the measurement is carried out (see Fig. 4.1).
The transducers are mounted on the outside of the pipe. They send and receive ultrasonic signals through the fluid.
The transmitter controls the measuring cycle, eliminates noise signals and analyzes useful signals. The measured values
can be displayed, used for calculations and transmitted.
Fig. 4.1:
Example of a measurement arrangement
1 – transducers
2 – supply line
3 – return line
4 – temperature probe on the supply line
5 – temperature probe on the return line
6 – transmitter
1
5
4
6
2
3