16 Advanced functions
16.9 Behavior of the alarm outputs
FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
120
The alarm state can be displayed after the configuration of the alarm outputs and during the measurement. This function is
activated in the menu item
Special Funct.\SYSTEM settings\Dialogs/Menus
. The activation of this function is
recommended when alarm outputs have to be reconfigured frequently.
• Select the menu item
SHOW RELAIS STAT
.
• Select
on
to activate the alarm state indication.
• Press ENTER.
If the alarm state indication is activated, the alarm output state is displayed after configuring the alarm outputs.
The alarm state indication is structured as follows:
Rx =
, with
x
being the number of the alarm output and
a pictogram according to Tab. 16.3.
The configuration of the alarm outputs can be repeated by pressing C. When the configuration of the alarm outputs is
finished, press ENTER. The main menu will be displayed.
If the alarm output indication is activated, the alarm state can be displayed during the measurement. Press
or
to
scroll through the upper or lower line, respectively, until the alarm state is displayed.
16.9.5
Deactivation of an alarm output
If the programmed outputs are no longer required, they can be deactivated. The configuration of a deactivated output is
stored and will be available if the output is activated again.
• Select
no
in the menu item
Output Options\Alarm Output
to deactivate an output.
• Press ENTER.
Special Funct.\SYSTEM settings\Dialogs/Menus\SHOW RELAIS STAT
Tab. 16.3: Pictograms for the alarm state indication
no.
func
(switching condition)
typ
(holding behavior)
mode
(switching function)
current state
R
=
OFF
NON-HOLD
NO Cont.
closed
1
MAX
HOLD
NC Cont.
open
2
MIN
3
+→- -→+
QUANT.
ERROR
Example
R1 =
Output Options\...\Alarm Output
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......