3 General principles
3.4 Undisturbed flow profile
FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
16
3.4
Undisturbed flow profile
Some flow elements (e.g., elbows, valves, pumps, reducers) distort the flow profile in their vicinity. The axisymmetrical
flow profile in the pipe needed for correct measurement is no longer given. A careful selection of the measuring point helps
to reduce the impact of disturbance sources.
It is most important that the measuring point is chosen at a sufficient distance from any disturbances. Only then it can be
assumed that the flow profile in the pipe is fully developed. However, measuring results can be obtained even if the
recommended distance to disturbances cannot be met for practical reasons (no ideal inflow, see section 16.4).
The recommended straight inlet and outlet pipe lengths for different types of flow disturbance sources are shown in the
examples in Tab. 3.1.
Fig. 3.11: Recommended transducer mounting position
Fig. 3.12: Disadvantageous transducer mounting position
Tab. 3.1:
Recommended distance from disturbance sources
D – nominal pipe diameter at the measuring point
l – recommended distance between disturbance source and transducer position
disturbance source: 90° elbow
inlet: l ≥ 10 D
outlet: l ≥ 5 D
disturbance source: 2 × 90° elbow on different planes
inlet: l ≥ 40 D
outlet: l ≥ 5 D
l
l
l
l
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......