7 Connection
FLUXUS F502TE
7.1 Transducers
55
UMFLUXUS_F502TEV1-5EN, 2020-05-29
7.1.5
Transducer module (SENSPROM)
The transducer module contains important transducer data for the operation of the transmitter with transducers.
If transducers are replaced, the sensor module must also be replaced.
• Disconnect the transmitter from the power supply.
• Insert the transducer module into the socket (see Fig. 7.9).
• Connect the transmitter to the power supply.
• Enter all parameters of the program branch
Parameter
.
• Start the measurement.
Notice!
The serial numbers of the transducer module and the transducer must be identical. A wrong or incorrectly connected
transducer module will lead to incorrect measured values or to a measurement failure.
Fig. 7.9:
Connection of the transducer module
1 – transducer module
AV
AV
S
AGN
ARS
AR
1