![Flexim FLUXUS F502BT Скачать руководство пользователя страница 83](http://html1.mh-extra.com/html/flexim/fluxus-f502bt/fluxus-f502bt_operating-instruction_2304725083.webp)
9 Measurement
FLUXUS F502TE
9.2 Measurement settings
83
UMFLUXUS_F502TEV1-5EN, 2020-05-29
• Press ENTER until
Current Loop
is displayed. Select
yes
to activate the output.
• Press ENTER.
Measuring range
After an analog output has been activated in the program branch
Output Options
, the measuring range of the source
item has to be entered.
• Select
sign
if the sign of the measured values is to be considered for the output.
• Select
absolut
if the sign of the measured values is not to be considered for the output.
• Press ENTER.
• Enter the lowest expected measured value. The unit of measurement of the source item will be displayed.
Zero-Scale Val.
is the value assigned to the value
Output MIN
of the output range.
• Press ENTER.
• Enter the highest expected measured value. The unit of measurement of the source item will be displayed.
Full-Scale Val.
is the value assigned to the value
Output MAX
of the output range.
• Press ENTER.
Function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the installed output.
• Select
yes
to test the output.
• Press ENTER.
• Enter a test value for the selected physical quantity. If the external measuring instrument displays the entered value, the
output functions correctly.
• Press ENTER.
Output Options\...\Current Loop
Output Options\...\Meas.Values
Output Options\...\Zero-Scale Val.
Output Options\...\Full-Scale Val.
Example
output: current output
output range: 4…20 mA
Zero-Scale Val.
: 0 m³/h
Full-Scale Val.
: 300 m³/h
volumetric flow rate = 0 m³/h, corresponds to 4 mA
volumetric flow rate = 300 m³/h, corresponds to 20 mA
Output Options\...\I1:Test output?
Output Options\...\I1:Test value
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......