10 Troubleshooting
10.6 Problems with the totalizers
FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
94
10.6
Problems with the totalizers
The values of the totalizers are too high.
See
Special Funct.\SYSTEM settings\Measuring\Quantity recall
. If this menu item is activated, the values
of the totalizers will be stored. The totalizers will continue with these values at the start of the next measurement.
The values of the totalizers are too small.
One of the totalizers has reached the upper limit and has to be reset to zero manually.
The sum of the totalizers is not correct.
See
Special Funct.\SYSTEM settings\Measuring\Quant. wrapping
. The sum of both totalizers (throughput)
transmitted via the output is no longer valid after the first overflow (wrapping) of one of the totalizers.
10.7
Problems during the thermal energy rate measurement
The measured fluid temperatures differ from the actual values.
The temperature probes are not sufficiently insulated.
On a pipe with a small diameter, the temperature probe is lifted from the pipe surface by the insulation foam.
The measured absolute value of the thermal energy rate is correct but has the opposite sign.
Check the assignment of the supply and return temperature to the temperature inputs.
The calculated thermal energy rate differs from the actual one although the measured flow and temperature
values are correct.
Check the thermal energy rate coefficients of the fluid.
Содержание FLUXUS F502BT
Страница 1: ...FLUXUS F502TE UMFLUXUS_F502TEV1 5EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F502TE 2020 05 29 UMFLUXUSF502TEV1 5EN 6...
Страница 144: ...19 Thermal energy rate measurement 19 4 Steam in inlet FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 144...
Страница 160: ...Annex D Conformity declarations FLUXUS F502TE 2020 05 29 UMFLUXUS_F502TEV1 5EN 160 D Conformity declarations...
Страница 161: ......
Страница 162: ......