11
EN
LARYNGOTEC
®
PRO TRACHEOSTOMY TUBES
I. FOREWORD
These instructions for use are valid for Fahl
®
tracheostomy tubes. The instructions for use serve to inform
the physician, nursing staff and patient/user in order to ensure technically correct handling of Fahl
®
tracheostomy tubes.
Please read the instructions for use carefully before using the product for the first time!
Keep them in an easily accessible place for future reference.
II. INTENDED USE
Fahl
®
tracheostomy tubes are intended to stabilise the tracheostoma following laryngectomy or
tracheostomy.
The tracheostomy tube is designed to keep the tracheostoma open.
When used for the first time the products must only be selected, used and inserted by clinicians or medical
professionals trained in their use.
Users must be trained by medical professionals on how to use the Fahl
®
tracheostomy tubes
safely and appropriately.
Fahl
®
LINGO tube variants are intended for tracheotomised patients who retain their larynx or for
laryngectomees who use a shunt valve (voice prosthesis users).
III. WARNINGS
Patients must be trained by medical professionals on how to use the Fahl
®
tracheostomy tubes
safely.
Fahl
®
tracheostomy tubes must never be blocked, e.g. by mucus or encrustation. Blockage can
cause asphyxiation!
Mucus in the trachea can be aspirated (removed by suction) through the tracheostomy tube with
a tracheal suction catheter.
Damaged tracheostomy tubes must not be used and must be discarded immediately. Use of a
damaged tube may result in airway compromise.
Do not use tracheostomy tubes during laser treatment (laser therapy) or with electrosurgical
devices! Contact with a laser beam may damage the tube.
When the tracheostomy tubes are inserted or removed, this can lead to irritations, coughing or
slight bleeding. If bleeding persists, consult your doctor immediately!
CAUTION!
Tracheostomy tubes with speech function are only recommended for tracheotomised patients
with normal secretion and normal mucous tissue.
CAUTION!
If there is strong secretion, a tendency to form granulation tissue, if the patient is undergoing
radiation therapy or if scabs have formed, a perforated cannula version is only recommended
under regular surveillance by a doctor and provided that shorter replacement intervals are
observed (as a rule once a week), because the perforations in the outer tube can increase the
formation of granulation tissue.
IV. COMPLICATIONS
The following complications could emerge when using this device:
Contamination of the stoma may make it necessary to remove the tube. Contaminations can furthermore
cause infections that necessitate the use of antibiotics.
Unintentional Inhalation of a tube/button that was not sized correctly may need to be removed by a doctor/
physician. If mucus blocks the tube/button it should be removed and cleaned.
V. CONTRAINDICATIONS
Do not use if the patient is allergic to the material.
Do not use with ventilated patients.
VI. PRECAUTIONS
The correct tracheostomy tube size should be selected by the treating physician or by trained medical
professionals.
When using Fahl
®
tracheostomy tubes, the products may become infected for instance with yeast
(Candida), bacteria or the like, which may lead to the material becoming cloudy and its service life being
shortened. Should this occur, the tube must be replaced immediately.
Use only auxiliary equipment with a 22 mm connection with Fahl
®
KOMBI adapters to prevent accidental
loss of accessories or damage to the tube.
We highly recommend always keeping at least two replacement tubes on hand to ensure continuous
supply.
Содержание LARYNGOTEC PRO
Страница 1: ...LARYNGOTEC PRO GEBRAUCHSANWEISUNG INSTRUCTIONS FOR USE...
Страница 2: ...2 DE BILDER PICTURES DE EN 4 1 5 7 3 9 10 8 2 6...
Страница 71: ...71 EL 7 6 A p 22 E s S l MRI J b I P n d Q 5 m...
Страница 72: ...72 EL LARYNGOTEC PRO I Fahl Fahl Fahl Fahl Fahl LINGO III Fahl Fahl IV V VI Fahl Candida...
Страница 75: ...75 EL IX Fahl MRSA ORSA 7 1 Fahl p 31110 Fahl 5 10 100 212 Fahrenheit 8 31200 9...
Страница 76: ...76 EL 8 OPTIBRUSH 31850 OPTIBRUSH Plus 31855 10 0 9 NaCl 2 Fahl OPTICIT 31180 0 9 NaCl X XI 6...
Страница 96: ...96 RU 7 6 A p 22 KOMBI E s S l J b I P n d Q 5 m...
Страница 97: ...97 RU LARYNGOTEC PRO I Fahl Fahl II Fahl Fahl Fahl LINGO III Fahl Fahl IV V VI Fahl...
Страница 132: ...132 BG 7 6 A p 22 mm KOMBI E s S l J b I P n d Q 5 m...
Страница 133: ...133 BG LARYNGOTEC PRO I Fahl Fahl II Fahl Fahl Fahl LINGO III Fahl Fahl p IV V VI Fahl Candida...
Страница 136: ...136 BG IX Fahl MRSA ORSA 7 1 Fahl pH REF 31110 Fahl 100 212 5 10 8 REF 31200 9 8...
Страница 137: ...137 BG OPTIBRUSH REF 31850 OPTIBRUSH Plus REF 31855 10 0 9 NaCl 2 Fahl OPTICIT REF 31180 Glutaraldehyd 0 9 NaCl X XI 6...
Страница 145: ...145 JA 7 6 A p 22mm KOMBI E s S l MRT J b I P n ID d Q 5 m...
Страница 146: ...146 JA LARYNGOTEC PRO I Fahl Fahl II Fahl Fahl Fahl LINGO III Fahl Fahl IV V VI Fahl Fahl 22mm...
Страница 155: ...155 AR 1 Fahl Fahl 2 Fahl Fahl Fahl LINGO 3 Fahl Fahl 4 5 6 Fahl Fahl 22 7 LARYNGOTEC Fahl Fahl LARYNGOTEC PRO...
Страница 156: ...156 AR 7 6 A 22 KOMBI p E s S l J b I P n d Q 5 m...
Страница 157: ...157 157...
Страница 162: ......
Страница 163: ......