9 Trouble-shooting
Proline Promag 51
90
Hauser
9.3
Process error messages
Process errors can be defined as either “Fault” or “Notice” messages and can thereby be weighted
differently. Determination of this is done via the function matrix (see the “Description of Device
Functions” Manual).
!
Note!
The error types listed in the following correspond to the factory settings.
Also observe the information on Page 56 ff. and 92
No.
Error message / Type
Cause
Remedy
P = Process error
$
= Fault message (with an effect on the outputs)
! = Notice message (without an effect on the outputs)
!
Note!
• In custody transfer mode, “notice messages” of Promag 51 are always classified and treated as “fault messages”.
• Error response of the outputs
401
P: EMPTY PIPE
$
: # 401
Measuring tube partially filled or
empty
1.
Check the process conditions of
the plant
2.
Fill the measuring tube
Custody transfer mode:
Reset error message via the status
input or via the “reset key”.
461
P: ADJ. NOT OK
!: # 461
EPD/OED adjustment not possible
because the fluid's conductivity is
either too low or too high.
The EPD/OED function cannot be
used with fluids of this nature.
463
FULL = EMPTY
$
: # 463
The EPD/OED adjustment values for
empty pipe and full pipe are identical,
therefore incorrect.
Repeat adjustment, making sure
procedure is correct
Custody transfer mode:
In custody transfer mode, an
EPD/OED adjustment is not possible.
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...