9 Trouble-shooting
Proline Promag 51
88
Hauser
359
…
362
S: PULSE RANGE n
!: # 359…362
Pulse output:
Pulse output frequency is out of
range.
1.
Increase the setting for pulse
weighting
2.
When selecting the pulse width,
choose a value that can still be
processed by a connected coun-
ter (e.g. mechanical counter,
PLC etc.).
Determine the pulse width:
– Variant 1: Enter the minimum
duration that a pulse must be
present at the connected
counter to ensure its registra-
tion.
– Variant 2: Enter the maximum
(pulse) frequency as the half
“reciprocal value” that a pulse
must be present at the con-
nected counter to ensure its
registration.
Example:
The maximum input frequency
of the connected counter is
10 Hz. The pulse width to be
entered is:
3.
Reduce flow.
Custody transfer mode:
If the pulse output is not suitable for
custody transfer mode, the pulse
value can be altered via the HART
interface or the ToF Tool - Fieldtool
Package.
The pulse width can be altered inde-
pendently of custody transfer mode.
No. # 5xx
→
Application error
501
S: SW.-UPDATE ACT.
!: # 501
New amplifier or communication
(I/O module) software version is
loaded. Currently no other functions
are possible.
Wait until the procedure is finished.
The device will restart automatically.
Custody transfer mode:
In custody transfer mode, no
software version can be loaded.
502
S: UP-/DOWNLOAD ACT
!: # 502
Up- or downloading the device data
via configuration program. Currently
no other functions are possible.
Wait until the procedure is finished.
Custody transfer mode:
In custody transfer mode, no up- or
downloading possible.
No. # 6xx
→
Simulation mode active
601
S: POS. ZERO RETURN
!: # 601
Positive zero return active
"
Caution!
This message has the highest display
priority!
Switch off positive zero return
Custody transfer mode:
In custody transfer mode, positive
zero return cannot be activated.
611
…
614
S: SIM. CURR. OUT. n
!: # 611...614
Simulation current output active
Switch off simulation
Custody transfer mode:
Simulation not possible
No.
Error message / Type
Cause
Remedy / spare part
1
2 10 Hz
⋅
----------------------
50 ms
=
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...