Connecting the measuring unit
4.2.1
Transmitter
#
Warning!
• Risk of electric shock. Switch off the power supply before opening the device.
Do not install or wire the device while it is connected to the power supply. Failure to comply with
this precaution can result in irreparable damage to the electronics.
• Risk of electric shock. Connect the protective conductor to the ground terminal on the housing
before the power supply is applied (not necessary if the power supply is galvanically isolated).
• Compare the specifications on the nameplate with the local voltage supply and frequency. The
national regulations governing the installation of electrical equipment also apply.
1.
Remove the cover of the connection compartment (f) from the transmitter housing.
2.
Feed the power supply cable (a) and signal cables (b) through the appropriate cable entries.
3.
Connect the cables in accordance with the wiring diagram:
– Wiring diagram (aluminium housing)
– Wiring diagram (wall-mount housing)
– Terminal assignment
4.
Screw the cover of the connection compartment (f) firmly onto the transmitter housing.
F06-xxxxxxxx-04-06-xx-xx-005
Fig. 29:
Connecting the transmitter (aluminium field housing). Cable cross-section: max. 2.5 mm
2
a
Cable for power supply: 85…260 V AC, 20…55 V AC, 16…62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L
−
for DC
b
Signal cable: Terminals Nos. 20–27
→
c
Ground terminal for protective conductor (Please also observe company-internal grounding concepts)
d
Ground terminal for signal cable shield
e
Service adapter for connecting service interface FXA 193 (Fieldcheck, ToF Tool - Fieldtool Package)
f
Cover of the connection compartment
g
Securing clamp
b
b
c
d
a
a
2
1
– 27
– 25
– 23
– 21
+ 26
+ 24
+ 22
+ 20
L1 (L+)
N (L-)
g
f
e
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...