Proline Promag 51
4 Wiring
Hauser
47
4.3
Potential equalisation
4.3.1
Standard case
Perfect measurement is only ensured when the medium and the sensor have the same electrical
potential. Most Promag sensors have a standard installed reference electrode which guarantees the
required connection. This usually means that no ground disks or additional potential matching
measures are necessary.
!
Note!
For installation in grounded metal pipes, it is advisable to connect the ground terminal of the trans-
mitter housing to the piping. Also, observe company-internal grounding guidelines.
F06-5xxxxxxx-04-xx-xx-xx-002
Fig. 34:
Potential equalisation by means of the transmitter's ground terminal
"
Caution!
For sensors without reference electrodes or without metal process terminals, carry out potential
equalisation as per the instructions for special cases described below. These special measures are
particularly important when standard grounding practice cannot be ensured or extremely strong
equalising
currents are expected.
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...