Proline Promag 51
10 Technical data
Hauser
109
"
Caution!
It is not allowed to use the device beyond the min. and max. lining specified temperature values
(
→
“Medium temperature range”).
Note the following points:
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• If both fluid and ambient temperatures are high, install the transmitter at a remote location from
the sensor (
→
“Medium temperature range”).
Storage temperature
The storage temperature corresponds to the operating temperature range of the measuring
transmitter and the appropriate measuring sensors.
Degree of protection
• Standard: IP 67 (NEMA 4X) for transmitter and sensor
• Optional: IP 68 (NEMA 6P) for remote version of Promag W and P sensor
Shock and vibration resistance
Acceleration up to 2 g by analogy with IEC 60068-2-6
(high-temperature version: no data available)
CIP cleaning
Promag W: not possible
Promag P: possible (note max. temperature)
SIP cleaning
Promag W: not possible
Promag P: possible with PFA (note max. temperature)
Electromagnetic
compatibility (EMC)
To EN 61326/A1 and NAMUR recommendation NE 21.
Process
Medium temperature range
Certified instruments:
0...+30 °C (cold water)
Certifiable instruments:
The permissible fluid temperature depends on the lining of the measuring tube:
Promag W:
0…+80 °C for hard rubber (DN 65…2000)
–20…+50 °C for polyurethane (DN 25…1000)
Promag P:
–40…+130 °C for PTFE (DN 15…600), for restrictions
→
refer to diagrams
–20…+180 °C for PFA (DN 25…200), for restrictions
→
refer to diagrams
Promag P
–40…+130 °C for PTFE (DN 15…600), for restrictions
→
refer to diagrams
–20…+180 °C for PFA (DN 25…200), for restrictions
→
refer to diagrams
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...