Proline Promag 51
10 Technical data
Hauser
107
10.1.5
Power supply
Electrical connections
Cable entry
Power supply and signal cables (inputs/outputs):
• Cable entry M20 x 1.5 (8…12 mm)
• Sensor cable entry for armoured cables M20 x 1.5 (9.5...16 mm) (not in custody transfer mode)
• Threads for cable entries 1/2" NPT, G 1/2"
Connecting cable for remote version:
• Cable entry M20 x 1.5 (8…12 mm)
• Sensor cable entry for armoured cables M20 x 1.5 (9.5...16 mm) (not in custody transfer mode)
• Threads for cable entries, 1/2" NPT, G 1/2"
Cable specifications
Supply voltage
85…260 V AC, 45…65 Hz
20…55 V AC, 45…65 Hz
16…62 V DC
Power consumption
AC: <15 VA (including sensor)
DC: <15 W (including sensor)
Switch-on current:
• max. 13.5 A (< 50 ms) at 24 V DC
• max. 3 A (< 5 ms) at 260 V AC
Power supply failure
Lasting min. 1 power cycle
• EEPROM saves measuring system data if power supply fails
• S-DAT: exchangeable data storage chip which stores the data of the sensor
(nominal diameter, serial number, calibration factor, zero point, etc.)
Potential equalisation
10.1.6
Performance characteristics
Reference operating
conditions
To DIN EN 29104 and VDI/VDE 2641:
• Fluid temperature: +28 °C ± 2 K
• Ambient temperature: +22 °C ± 2 K
• Warm-up time: 30 minutes
Installation:
• Inlet run >10 x DN
• Outlet run > 5 x DN
• Sensor and transmitter grounded.
• Sensor centered relative to the pipe.
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...