10 Technical data
Proline Promag 51
108
Hauser
Maximum measured error
Pulse output:
• ± 0.5% o.r. ± 1 mm/s (o.r. = of reading)
• Option: ± 0.2% o.r. ± 2 mm/s (o.r. = of reading)
Current output: plus typically ± 5
µ
A
Supply voltage fluctuations have no effect within the specified range.
F06-5xxxxxxx-05-xx-xx-xx-000
Fig. 50:
Max. measured error in % of reading
Repeatability
max. ± 0.1% o.r. ± 0.5 mm/s (o.r. = of reading)
10.1.7
Operating conditions
Installation
Installation instructions
Any orientation (vertical, horizontal)
Restrictions and additional installation instructions
→
Inlet and outlet runs
Inlet run: typically
≥
5 x DN
Outlet run: typically
≥
2 x DN
Custody transfer mode: see remarks on Page 20
Length of connecting cable
For the remote version the permissible cable length L
max
depends on the conductivity
of the medium
A minimum conductivity of 20
µ
S/cm is required for measuring demineralized water.
Environment
Ambient temperature
Transmitter:
• Standard: –20...+60 °C (Custody transfer mode)
• Optional: –40...+60 °C (Usual transfer mode)
!
Note!
At ambient temperatures below –20 °C the readability of the display may be impaired.
(not in custody transfer mode)
Sensor:
• Flange material carbon steel: –10...+60 °C
• Flange material stainless steel: –40...+60 °C
0
1
2.5
[%]
2.0
1.5
1.0
0.5
0
2
4
6
8
10
v [m/s]
0.2 %
0.5 %
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...