Assembly with ground disks (DN 15…300)
Depending on the application, e.g. with lined or ungrounded pipes (see Page 47 ff.), it may be nec-
essary to mount ground disks between the sensor and the pipe flange for the potential equalisation.
Ground disks can be ordered separately as an accessory from Hauser (see Page 81).
"
Caution!
• In this case, when using ground disks (including seals) the total fitting length increases!
All the dimensions are provided in the separate documentation “Technical Information”.
• PTFE and PFA lining
→
only install additional seals between the ground disk and pipe flange.
1.
Place the ground disk and the additional seal between the instrument and the pipe flange
(Fig. 21).
2.
Insert the bolts through the flange holes. Tighten the nuts so that they are still loose.
3.
Now rotate the ground disk as shown in Fig. 21 until the handle strikes the bolts.
This will center the ground disk automatically.
4.
Now tighten the bolts to the required torque (see Page 34 ff.)
5.
Connect the ground disk to ground
F06-5xFxxxxx-17-05-xx-xx-001
Fig. 21:
Assembly with ground disks (Promag P, DN 15…300)
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...