Proline Promag 51
3 Installation
Hauser
35
3.3.3
Turning the transmitter housing
#
Warning!
The turning mechanism in devices with EEx d/de or FM/CSA Cl. I Div. 1 classification is not the
same as that described here. The procedure for turning these housings is described in the Ex-specific
documentation.
1.
Loosen the two securing screws.
2.
Turn the bayonet catch as far as it will go.
3.
Carefully lift the transmitter housing as far as it will go.
4.
Turn the transmitter housing to the desired position (max. 2 x 90° in either direction).
5.
Lower the housing into position and re-engage the bayonet catch.
6.
Retighten the two securing screws.
F06-xxxxxxxx-17-06-xx-xx-000
Fig. 22:
Turning the transmitter housing (aluminum field housing)
3.3.4
Turning the local display
1.
Remove the cover of the electronics compartment.
2.
Press the side latches on the display module and remove it from the electronics compartment
cover plate.
3.
Rotate the display to the desired position (max. 4 x 45° in each direction), and place it back
into the electronics compartment cover plate.
4.
Screw the cover of the electronics compartment firmly onto the transmitter housing.
F06-xxxxxxxx-07-xx-06-xx-000
Fig. 23:
Turning the local display (field housing)
3
5
6
1
2
4
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...