Proline Promag 51
5 Operation
Hauser
55
5.2.1
General notes
The Quick Setup menu (see Page 76) is adequate for commissioning in most instances. Complex
measuring operations on the other hand necessitate additional functions that you can configure as
necessary and customize to suit your process parameters. The function matrix, therefore, comprises
a multiplicity of additional functions which, for the sake of clarity, are arranged in a number of func-
tion groups.
Comply with the following instructions when configuring functions:
• You select functions as described on Page 54.
• You can switch off certain functions (OFF). If you do so, related functions in other
function groups will no longer be displayed.
• Certain functions prompt you to confirm your data entries. Press
OS
to select “SURE [ YES ]”
and press
F
again to confirm. This saves your setting or starts a function, as applicable.
• Return to the HOME position is automatic if no key is pressed for 5 minutes.
!
Note!
• The transmitter continues to measure while data entry is in progress, i.e. the current measured
values are output via the signal outputs in the normal way.
• If the power supply fails, all preset and parameterized values remain safely stored in the EEPROM.
"
Caution!
All functions are described in detail, including the function matrix itself, in the
“Description of
Device Functions”
manual, which is a separate part of this Operating Instruction.
5.2.2
Enabling the programming mode
The function matrix can be disabled. Disabling the function matrix rules out the possibility of
inadvertent changes to device functions, numerical values or factory settings.
A numerical code (factory setting = 51) has to be entered before settings can be changed.
If you use a code number of your choice, you exclude the possibility of unauthorized persons
accessing data (
→
see the “Description of Device Functions” manual).
Comply with the following instructions when entering codes:
• If programming is disabled and the
OS
key is pressed in any function, a prompt for the code
automatically appears on the display.
• If “0” is entered as the customer's code, programming is always enabled.
• The Hauser service organisation can be of assistance if you mislay your personal code.
"
Caution!
Changing certain parameters such as all sensor characteristics, for example, influences numerous
functions of the entire measuring system, particularly measuring accuracy.
There is no need to change these parameters under normal circumstances and consequently, they
are protected by a special code known only to the Hauser service organization. Please con-
tact Hauser if you have any questions.
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...