6 Commissioning
Proline Promag 51
76
Hauser
6.4
Application-specific commissioning
In the case of measuring devices without a local display, the individual parameters and functions
must be configured via the configuration program, e.g. ToF Tool - Fieldtool Package. If the
measuring device is equipped with a local display, all the important device parameters for standard
operation can be configured quickly and easily by means of the “Commissioning” Quick Setup
menu.
6.4.1
“Commissioning” Quick Setup menu
This Quick Setup menu guides you systematically through the setup procedure for all the major
device functions that have to be configured for standard measuring operation.
!
Note!
In the case of measuring devices
without
a local display, the individual parameters and functions
must be configured by means of a configuration program, such as ToF Tool - Fieldtool Package from
Hauser.
F06-50xxxxxx-19-xx-xx-en-000
Fig. 43:
Quick Setup for commissioning
+
+
+
E
E
-
+
HAUSER
E
ESC
QS
Commission
Language
Defaults
Quick Setup
Unit
Volume flow
Measuring
Mode
HOME-POSITION
Frequency
Pulse
Current Output
Freq.-/ Pulse Output
Quit
Assign
Current
Assign
Frequency
Assign
Pulse
Current
Span
Value
20 mA
Time
Constant
Time
Constant
Failsafe
Mode
Failsafe
Mode
Failsafe
Mode
Operation
Mode
End Value
Freq.
Value
f max
Pulse
Value
Pulse
Width
Output
Signal
Output
Signal
Automatic parameterization
of the display
Quit Quick Setup
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...