Dosimag
Commissioning
Hauser
37
Parameter
Prerequsite
Description
Selection / User
entry
Factory setting
Failure frequency
In the
Operating mode
Frequency
option is selected
and in the
Assign frequency
output
parameter (→ 36),
the
Volume flow
option is
selected.
Enter frequency output value
in alarm condition.
0.0 to 10
000.0 Hz
0.0 Hz
Invert output signal
–
Invert the output signal.
• No
• Yes
• Pulse/freq./switch
output 1: yes
• Pulse/freq./switch
output 2: no
Configuring the switch output
Navigation
"Setup" menu → Pulse/frequency/switch output 1 to 2
Structure of the submenu for the switch output
‣
Pulse/frequency/switch output 1
to 2
Operating mode
Switch output function
Assign diagnostic behavior
Assign limit
Assign flow direction check
Assign status
Switch-on value
Switch-off value
Failure mode
Invert output signal