Commissioning
Dosimag
38
Hauser
Parameter overview with brief description
Parameter
Prerequsite
Description
Selection / User
entry
Factory setting
Operating mode
–
Define the output as a pulse,
frequency or switch output.
• Off
• Pulse
• Automatic pulse
• Frequency
• Switch
•
Pulse/frequency/
switch output 1
submenu:
Pulse
option
•
Pulse/frequency/
switch output 1 to
2
submenu:
Switch
option
Switch output function
The
Switch
option is selected
in the
Operating mode
parameter.
Select function for switch
output.
• Off
• On
• Diagnostic
behavior
• Limit
• Flow direction
check
• Status
Off
Assign diagnostic behavior
• The
Switch
option is
selected in the
Operating
mode
parameter.
• The
Diagnostic behavior
option is selected in the
Switch output function
parameter.
Select diagnostic behavior for
switch output.
• Alarm
• Alarm or warning
• Warning
Alarm
Assign limit
• The
Switch
option is
selected in the
Operating
mode
parameter.
• The
Limit
option is selected
in the
Switch output
function
parameter.
Select process variable for limit
function.
Volume flow
Volume flow
Assign flow direction check
• The
Switch
option is
selected in the
Operating
mode
parameter.
• The
Flow direction check
option is selected in the
Switch output function
parameter.
Select process variable for flow
direction monitoring.
• Off
• Volume flow
Volume flow
Assign status
• The
Switch
option is
selected in the
Operating
mode
parameter.
• The
Status
option is
selected in the
Switch
output function
parameter.
Select device status for switch
output.
Low flow cut off
Low flow cut off
Switch-on value
• The
Switch
option is
selected in the
Operating
mode
parameter.
• The
Limit
option is selected
in the
Switch output
function
parameter.
Enter measured value for the
switch-on point.
Signed floating-point
number
Country-specific:
• 0 ml/s
• 0 fl oz/s
Switch-off value
• The
Switch
option is
selected in the
Operating
mode
parameter.
• The
Limit
option is selected
in the
Switch output
function
parameter.
Enter measured value for the
switch-off point.
Signed floating-point
number
Country-specific:
• 0 ml/s
• 0 fl oz/s