Diagnostics and troubleshooting
Hauser
47
Diagnostic information
The fault can be identified using the diagnostic information. The short text helps you by
providing information about the fault.
Diagnostic information
Diagnostic code
Status signal
Diagnostic number
Short text
↓
↓
↓
Example
A0013958
842
Process limit
3-digit number
12.2.2 Calling up remedy information
Remedy information is provided for every diagnostic event to ensure that problems can be
rectified quickly:
• On the home page
Remedy information is displayed in a separate field below the diagnostics information.
• In the
Diagnostics
menu
Remedy information can be called up in the working area of the user interface.
The user is in the
Diagnostics
menu.
1. Call up the desired parameter.
2. On the right in the working area, mouse over the parameter.
A tool tip with remedy information for the diagnostic event appears.
12.3 Adapting the diagnostic information
12.3.1 Adapting the diagnostic behavior
Each item of diagnostic information is assigned a specific diagnostic behavior at the
factory. The user can change this assignment for certain diagnostic information in the
Diagnostic behavior
submenu.
"Expert" menu → System → Diagnostic handling → Diagnostic behavior
You can assign the following options to the diagnostic number as the diagnostic behavior:
Options
Description
Alarm
Measurement is interrupted. Signal outputs and totalizers assume the defined alarm
condition. A diagnostic message is generated.
Warning
Measurement is resumed. The signal outputs and totalizers are not affected. A
diagnostic message is generated.
Logbook entry only
The device continues to measure. The diagnostic message is entered in the Event
logbook (events list) submenu only and is not displayed in alternation with the measured
value display.
Off
The diagnostic event is ignored, and no diagnostic message is generated or entered.
12.4 Overview of diagnostic information
In the case of some items of diagnostic information, the status signal and the
diagnostic behavior can be changed. Change the diagnostic information → 47